Humidity
Table of contents
- How is the ordering code "cable length" defined?
- Is it possible to operate a transmitter with a 2-wire-output (output: 4...20mA) with AC supply?
- What is a coating-procedure exactly?
- When is a high-humidity calibration recommended?
- Can E+E's industrial sensing probes be operated at 0 bar (vacuum)?
- Can E+E's industrial sensing probes be operated at 25 bar?
- Is it possible to combine the switching outputs (option SW) with 2 connectors for supply/outputs and RS232 interface (option C07)?
- Can the configuration of the outputs (EE29/31/35/36 series) be changed in the field?
- Is the EE31 transmitter sending clear text messages to the hyper-terminal?
- How is it possible to query data of an EE31 via a serial connection?
- How much is the maximum allowed velocity of the pressure-tight sensing probe of the EE31?
- Can the EE31 transmitters be employed for dew point measurements as well?
- Can customers calibrate the humidity of the EE21?
- Can customers calibrate the temperature of the EE21?
- Is the humidity measurement of the EE21 without temperature output also temperature compensated?
- Is it possible to calibrate temperature as well with the humidity calibrator HUMOR 20?
- Which calibration intervals are recommended?
How is the ordering code "cable length" defined?
The ordering code "cable length" is made up of the ordering code "probe length" and the real cable length.e.g. your choice was ordering code "probe length" = 200mm and ordering code "cable length" = 5m -> then it results in a real cable length of 5m "cable length" - 200mm "probe length" = 4.8m real cable length
Is it possible to operate a transmitter with a 2-wire-output (output: 4...20mA) with AC supply?
No, that is impossible.What is a coating-procedure exactly?
This is an additional humidity permeable protective film, which is applied directly onto the humidity sensor.Advantage: better resistance of the sensor against corrosion, parallel resistors through conductive particles (low-ohm connections) are avoided
Disadvantage: slightly longer response-time