Temperature Calibration

The E+E accredited laboratory performs calibrations of temperature measuring devices in air for the range -70 to 200 °C (-94 to 392 °F) as well as in dry block calibrators for the range -45 to 425 °C (-49 to 797 °F). The E+E on-site service is available for accurate calibration of stationary systems such as furnaces, climate chambers, refrigeration and freezer cabinets.

Scope of Accreditation E+E Calibration Laboratory

Calibration Calibration object Measurement conditions Measurement range Calibration uncertainty
AA 0608 Lab Thermometer for
air temperature
comparison measurement
in temperature controlled
measuring chamber
-80...200 °C 0.03 K
AA 0608 Lab Contact thermometer
(Immersion- and cut-in probe)
comparison measurement
in dry block calibrator
-45...425 °C <23 °C: 0.5 * |t - 23| + 28 mK
≥23 °C: 0.22 * |t - 23| + 28 mK

 

Calibration Procedure

The calibration is performed for air temperature sensors in a temperature-stabilised measurement chamber at a gas flow of 5 l/min, for contact thermometer in a block calibrator with high precision temperature regulation. The generated temperature is determined using a calibrated Pt100 resistance thermometer and used as a reference value.

Inquire Calibration

Accompanying Slip

For efficient processing of your calibration order, please also send us a fully completed accompanying slip with the measuring device to be checked. 

Download here  >>

Downloads Temperature

LinkedIn Logo