EE871 operates on the dual wavelength NDIR principle, which compensates for ageing effects, is highly insensitive to pollution and stands for outstanding long-term stability even in demanding environment.

A multiple point CO2 and temperature adjustment procedure leads to excellent CO2 measurement accuracy over the entire temperature working range, ideal for use in agriculture or outdoors for instance. The measured data range of up to 5 % CO2 (50,000 ppm) is available on E2 digital interface and up to 1 % CO2 (10,000 ppm) is available on Modbus RTU interface.

The IP65 enclosure of probe and the replaceable filter offer excellent protection in harsh, polluted environment. With a special filter cap, the probe can be employed in applications with periodical H2O2 sterilization. The compact size, the M12 connector and the optional mounting flange allow for fast probe installation, replacement or removal during the cleaning of the site, for instance a stable or an incubator.

An optional kit facilitates easy configuration and adjustment of EE871. The measurement interval can be set according to the application requirements, by this the average current consumption can be reduced to 120μA for battery-operated devices.

技术资料

  • 0...2000 ppm
    0...5000 ppm
    0...10000 ppm

    测量范围
  • 精度±(50ppm +测量值的2%)§±(50ppm +测量值的3% )§±(100ppm +测量值的5% )
  • Modbus / E2 数字接口

    输出
  • 4.75-7.5V DC