高端温湿度变送器,工作温度达180°C

Open Link in Lightbox
  • Service
  • Humidity
  • Temperature

EE310

经过优化的EE310可在苛刻的工业应用中进行可靠的测量。除了精密测量相对湿度和温度以外,变送器还可以计算露点,绝对湿度和混合比等参数。

  • EE310 湿度/温度 带显示变送器

  • EE310 湿度/温度分体探头型变送器

  • EE310 湿度/温度墙面安装型变送器

  • EE310 温湿度变送器不锈钢外壳

  • EE310 UL认证版本

  • NEW! Now with Extended Warranty Option

    多种安装方式,包括墙面,管道和分体探头等。分体探头可工作于最高至180 °C (356 °F)温度的环境中,并且配备耐压接头的探头最高工作压力可至20bar(290 psi)。坚固耐用的聚碳酸酯外壳的设计更方便了安装和设备维护。通过两路模拟信号或Modbus RTU数字接口可读取测量值。

    先进的TFT彩色显示屏可最多同时显示四个测量值,并提供多种错误诊断。集成的数据记录功能保存所有的测量值和计算值。记录的数据直接以图表的形势显示在屏幕上并且可以简单的通过USB接口下载。

    E+E创新的涂层技术可保护敏感元件免受腐蚀性和导电性污染。直接通过显示屏或者通过USB接口由免费的PCS10软件进行产品配置和温湿度校准。

    典型应用

    • 工业过程检测及控制
    • 食品和药品
    • 工业干燥和加湿器
    • 气象和环境箱
    • 洁净室

    EXTEND YOUR WARRANTY UP TO 5 YEARS WITH THE NEW 
    E+E CALIBRATION SERVICE PACKAGES

    Our calibration service packages ensure that you are on the safe side when it comes to the accuracy and vailability of your measuring devices. Choose the right service package for your E+E product and benefit rom professional device service, calculable costs and excellent measuring accuracy in the long-term. Here you can find more detailed information.

    NEW! Now with Extended Warranty Option

    特征

    • 高精度测量
    • 工作温度范围广
    • 可计算多种湿度相关物理量
    • 带按钮的TFT彩屏
    • 可保存20000个测试值的集成数据记录器
    • 屏幕或LED显示错误自诊断
    • 听觉和视觉错误警告
    • 两路模拟量输出和MODBUS RTU数字接口
    • 坚固的聚碳酸酯IP65防护等级外壳
    • 模块化外壳概念,便于安装维护
    • 耐压接头最高工作压力至20bar(290 psi)
    • 插拔式探头更易于快速更换和安装
    • 100...240 VAC 电源模块
    • 创新的涂层技术可有效防止环境对元件污染
    • 通过显示屏或软件可简单的现场校准和配置

    技术资料

    测量范围
    0...100% RH
    -80...180 °C
    精度
    ±1.3% RH
    ±0.2 °C
    数字接口
    RS485带Modbus RTU接口
    输出
    0-1/5/10V, 0/4-20mA
    供电电源
    24V DC

    Calibrating your limits

    您的测量工具需要和您的产品和服务一样保持准确度。E+E Elektronik 的校准团队拥有 20 多年的专业经验,致力于为您提供可靠的服务,有效防止出现延误或偏差现象。

    发现与校准机构的合作价值,您需要优先考虑

    • 准确性、
    • 速度
    • 可靠性

    了解更多

    下载 EE310

    FAQ

    The device requires a firmware update. Connect the device via the USB service interface to a PC running PCS10 Product Configuration Software and confirm the update to version 1.20 or greater.

    Use the RS485 kit for retrofitting (order number HA010605), see the data sheet and operation manual of your specific E+E product.

    The device indicates an error according to NAMUR NE43. For details, please see the user manual of your specific E+E product.

    The device has one GND connection for the power supply, an analogue output GND and an interface GND, see connection diagram in the user manual. The analogue output GND and the interface GND carry only the current output resp. interface current, which results in a more stable signal compared to a mix GND connection.

    Press 2x ► | 5x ◄ | 1x ▲ | 3x ◄ | 1x ▼ | 1x ► then select the "Configuration menu" and press ENTER. Scroll down to “Diagnosis”, press ENTER and scroll down to “Password protection” here you can see the previously set password.

    Connect the device via the USB service interface to a PC running PCS10 Product Configuration Software and read out the communication setting.

    Alternative for device with display: the display configuration menu has a section for Modbus settings (RS485, Modbus RTU) or IP settings (Ethernet, Modbus TCP) where you can see the actual setting or change it.

    Connect the device via the USB service interface to a PC running PCS10 Product Configuration Software and see at the section "Log data download" for the data memory status.

    Remote probe types can be retrofitted with an Ethernet interface using HA010606, see data sheet and operation manual of your specific E+E product.

    The measured data logged by the device can only be downloaded via the USB service interface to a PC running the PCS10 Product Configuration Software.

    The pluggable remote probe has a label with the serial number close to the connector. This must be the same as the serial number of the basis device. Refer to the operation manual further details.

    The main difference is the hardware interface. The Modbus RTU protocol runs on RS485 Hardware while Modbus TCP/IP protocol runs on Ethernet hardware.