Humidity Calibration Kit

For a quick humidity check

The humidity transmitters as all other measuring instruments shall be periodically checked and eventually adjusted. The most simple solution for this purpose is the E+E Humidity Calibration Kit.

Inquire now

The E+E Humidity Calibration Kit is a cost effective solution for calibrating humidity measuring instruments with sensing probes Ø 10-12 mm (0.4-0.47 inch). It is very easy to use and it does not require highly qualified technical staff. The kit consists of a humidity calibration chamber and a choice of E+E Humidity Standard Sets. 

Humidity Standards

The E+E Humidity Standards are non-saturated salt solutions available in sets of five or fifty single-use ampoules, which may be stored an indefinite time. The salt solutions are non-harmful, handling them does not require specific safety measures. Safety data sheet is available upon request. Each E+E Humidity Standard Set is supplied with a traceable calibration certificate, issued by the Austrian National Metrology Institute (NMI).

You want to know more about our products?
We are looking forward to help you.

Technical data Humidity Calibration Kit

  • 0% RH ... ±0.3% RH
    5% RH ... ±0.5% RH
    10% RH ... ±0.5% RH
    20% RH ... ±0.5% RH
    35% RH ... ±0.5% RH

    humidity value in %RH ... accuracy at 23 °C (73.4 °F)
  • 50% RH ... ±0.9% RH
    65% RH ... ±0.9% RH
    80% RH ... ±1.2% RH
    95% RH ... ±1.2% RH

    humidity value in %RH ... accuracy at 23 °C (73.4 °F)