Temperature Calibration

The E+E accredited laboratory performs calibrations of temperature measuring devices in air for the range -80 to 200 °C (-112 to 392 °F) as well as in dry block calibrators for the range -45 to 425 °C (-49 to 797 °F) and in wind tunnels for the range from 15 °C to 50 °C under a fl ow of 1-40 m/s.
The E+E on-site service is available for accurate calibration of stationary systems such as furnaces, climate chambers, refrigeration and freezer cabinets.

Scope of Accreditation E+E Calibration Laboratory

Calibration Calibration object Measurement conditions Measurement range Calibration uncertainty
AA 0608 Lab Thermometer for
air temperature
comparison measurement in temperature controlled measuring chamber -80...200 °C 0.03 K
AA 0608 Lab Contact thermometer
(Immersion- and cut-in probe)
comparison measurement in dry block calibrator -45...425 °C <23 °C: 0.5 * |t - 23| + 28 mK
≥23 °C: 0.22 * |t - 23| + 28 mK
AA 0608 Lab Contact thermometer
(Immersion- and cut-in probe)
comparison measurement in wind tunnel under fl ow (1-40 m/s) 15...50 °C v=(1...<2) m/s: 522 mK
v=(2...40) m/s: 444 mK

 

Calibration Procedure

The calibration is performed for air temperature sensors in a temperature-stabilised measurement chamber at a gas flow of 5 l/min, for contact thermometer in a block calibrator with high precision temperature regulation. The generated temperature is determined using a calibrated Pt100 resistance thermometer and used as a reference value.

Inquire Calibration

Accompanying Slip

For efficient processing of your calibration order, please also send us a fully completed accompanying slip with the measuring device to be checked. 

Download here  >>

Downloads Temperature

LinkedIn Logo

Live Chat

This website is using Tidio, a chat platform that connects users with the customer support of E+E Elektronik Ges.m.b.H. We are collecting email addresses/names/phone numbers only with the consent of the users, in order to start the chat. The messages and data exchanged are stored within the Tidio application. For more information, please refer to their Privacy Policy. E+E Elektronik Ges.m.b.H is not making use of these messages or data other than to follow up on users’ registered issues or inquiries. Your personal data will be processed and transmitted in accordance with the General Data Protection Regulation (GDPR).