Temperature Calibration

The E+E accredited laboratory performs calibrations of temperature measuring devices in air for the range -80 to 200 °C (-112 to 392 °F) as well as in dry block calibrators for the range -45 to 425 °C (-49 to 797 °F) and in wind tunnels for the range from 15 °C to 50 °C under a fl ow of 1-40 m/s.
The E+E on-site service is available for accurate calibration of stationary systems such as furnaces, climate chambers, refrigeration and freezer cabinets.

Scope of Accreditation E+E Calibration Laboratory

Calibration Calibration object Measurement conditions Measurement range Calibration uncertainty
AA 0608 Lab Thermometer for
air temperature
comparison measurement in temperature controlled measuring chamber -80...200 °C 0.03 K
AA 0608 Lab Contact thermometer
(Immersion- and cut-in probe)
comparison measurement in dry block calibrator -45...425 °C <23 °C: 0.5 * |t - 23| + 28 mK
≥23 °C: 0.22 * |t - 23| + 28 mK
AA 0608 Lab Contact thermometer
(Immersion- and cut-in probe)
comparison measurement in wind tunnel under fl ow (1-40 m/s) 15...50 °C v=(1...<2) m/s: 522 mK
v=(2...40) m/s: 444 mK

 

Calibration Procedure

The calibration is performed for air temperature sensors in a temperature-stabilised measurement chamber at a gas flow of 5 l/min, for contact thermometer in a block calibrator with high precision temperature regulation. The generated temperature is determined using a calibrated Pt100 resistance thermometer and used as a reference value.