Pressure Calibration

E+E accredited laboratory performs calibrations of measuring devices for absolute, for positive and negative gauge pressure as well as for differential pressure:

Absolute pressure: 0.1…101 bar (1.45...1465 psi)
Relative pressure: -0.9…100 bar (-13...1450 psi)
Differential pressure: -0.3...9 bar (-43.5...131 psi)

Scope of Accreditation E+E Calibration Laboratory

Calibration Calibration object Measurement conditions Measurement range Calibration uncertainty
AA 0608 Lab Absolute pressure sensor (23 ±3) °C 750...1000 mbar
1000...1250 mbar
0.0875 mbar - 0.00132 % of m.v.
0.0611 mbar + 0.00132 % of m.v.
AA 0608 Lab Absolute pressure sensor (23 ±3) °C 0.1...1 bar
1...21 bar
21...101 bar
0.00028 bar
0.01 % of m.v. + 0.00018 bar
0.011 % of m.v. + 0.00006 bar
AA 0608 Lab Relative pressure sensor (23 ±3) °C -0.9...0 bar
0...<20 bar
20...100 bar
0.00028 bar
0.01 % of m.v. + 0.00028 bar
0.011 % of m.v. + 0.00016 bar
AA 0608 Lab Differential pressure sensor (23 ±3) °C 0...0.3 bar
0.3...9 bar
0.19 % of m.v. + 0.00005 bar
0.16 % of m.v. + 0.00006 bar
AA 0608 Lab Differential pressure sensor (23 ±3) °C -0.3...0.3 bar
-0.05...0.05 bar
0.0000342 bar + 0.00000274 % of m.v.
0.0000066 bar + 0.0000024 % of m.v.

 

Calibration Procedure

A pressure calibrator is used for the calibration of absolute and relative pressure meters. The generator consists of a pressure regulating unit, as well as two quartz relative pressure sensors for reference, and a quartz barometer.

The pressure drop of a gas stream is measured at a needle valve for the calibration of differential pressures. 

Inquire Calibration

Accompanying Slip

For efficient processing of your calibration order, please also send us a fully completed accompanying slip with the measuring device to be checked. 

Download here  >>

Downloads Pressure

LinkedIn Logo

Live Chat

This website is using Tidio, a chat platform that connects users with the customer support of E+E Elektronik Ges.m.b.H. We are collecting email addresses/names/phone numbers only with the consent of the users, in order to start the chat. The messages and data exchanged are stored within the Tidio application. For more information, please refer to their Privacy Policy. E+E Elektronik Ges.m.b.H is not making use of these messages or data other than to follow up on users’ registered issues or inquiries. Your personal data will be processed and transmitted in accordance with the General Data Protection Regulation (GDPR).