Humidity Calibration

The E+E calibration laboratory is accredited for relative humidity calibrations in the temperature range -70 °C to 200 °C (-94 °F to 392 °F). The reference is a high-end, sophisticated dual pressure – dual temperature humidity generator provides highest accuracy for the calibration of hygrometers and humidity generators. E+E Elektronik is also a Designated Institute (DI) responsible for maintaining the national standard for humidity in Austria.

Scope of Accreditation E+E Calibration Laboratory

Calibration Calibration object Measurement conditions Measurement range1) Calibration uncertainty
NMI Lab Special calibration of air velocity measuring instruments in the Designated Institute BEV/E+E
AA 0608 Lab Hygrometer for the measurement of humidity Gas flow max. 5 l / min
Air pressure max. 10 bar
Temperature range -40...180 C°1)

0...98 % RH; (-70...-20) °C
0...98 % RH; (-20...200) °C

(0.1 % * 4 * 10-6* (-20 °C-t)3 + (0.002 - 10-5*t) * Uw
0.1 % + (0.002 - 10-5 * t) *Uw
AA 0608 Lab Humidity generator (e.g. Humor 20)2) Temperature range (25 ±3) °C -20...0 °C
>0 °C; 1...10) % RH
>0 °C; 10...95) % RH
0.1 % RH + 0.007 % * Uw
0.1 % RH
0.05 % RH + 0.005 % * Uw

1) Calibration in the extended temperature range -70 °C to 200 °C (-94 °F to 392 °F) is performed upon request. Calibration for other physical quantities such as absolute humidity or mixing ratio is available upon request.
2) Wider ranges or dew point upon request.

Calibration Procedure

Air is generated (displayed) with a defined dew point temperature and a maximum gas flow of 5 l/min in a dual pressure, dual temperature humidity generator. This air stream is passed through a temperature-stabilised measurement chamber and there it produces a defined air humidity level, depending on the temperature, pressure and the displayed dew point temperature.

The humidity generated at the measurement location is determined by measuring the dew point temperature by means of a reference dew point mirror and the temperature by means of a calibrated Pt100 resistance thermometer and used as a reference value.

Inquire Calibration

Accompanying Slip

For efficient processing of your calibration order, please also send us a fully completed accompanying slip with the measuring device to be checked. 

Download here  >>

Downloads Humidity

LinkedIn Logo

Live Chat

This website is using Tidio, a chat platform that connects users with the customer support of E+E Elektronik Ges.m.b.H. We are collecting email addresses/names/phone numbers only with the consent of the users, in order to start the chat. The messages and data exchanged are stored within the Tidio application. For more information, please refer to their Privacy Policy. E+E Elektronik Ges.m.b.H is not making use of these messages or data other than to follow up on users’ registered issues or inquiries. Your personal data will be processed and transmitted in accordance with the General Data Protection Regulation (GDPR).