Flow Calibration

The E+E accredited calibration bench is unique in Europe and offers flow and volumetric flow calibrations for inline and immersion flow meters at standard conditions in the range from 0.6 to 2300 m³/h (0.35 to 1354 SCFM) in pipes from DN15 to DN80 (1/2 to 3“). The pressure in the calibration bench can be set between 1 and 10 bar (14.5 and 145 psi), which allows for highly accurate calibration under the real conditions in compressed air networks.

The E+E calibration laboratory operates also an accredited low flow calibration bench in the range from 0.06 l/min to 245 l/min at standard conditions. Calibrations can be performed in pipes from DN15 to DN50, at pressures between 1 and 10 bar (14.5 and 145 psi) and for air and non-air gases.

Scope of Accreditation E+E Calibration Laboratory

Calibration Calibration object Measurement conditions Measurement range Calibration uncertainty
AA 0608 Lab Flow sensor 0.1...1 MPa (1...10 bar)
at (23 ±3) °C
Flow rate at standard conditions (0 °C / 1013 mbar)
0.6...2300 m3/h
0.003 m3/h + 0.9 % of m.v.
AA 0608 Lab Flow sensor 0.1...1 MPa (1...10 bar)
at (23 ±3) °C
0.06...245 l/min 1) (0.12 * Q02 - 0.48 * Q0 + 1.66) % (Range: 0.06...2.7 l/min)
(0.008 * Q02 - 0.14 * Q0 + 1.86) % (Range: 1.9...9.3 l/min)
(0.0007 * Q02 - 0.052 * Q0 + 2.4) % (Range: 9...46 l/min)
(0.00004 * Q02 - 0.0018 * Q0 + 1.8) % (Range: 45...245 l/min)

1) also for non-air gases

Inquire Calibration

Accompanying Slip

For efficient processing of your calibration order, please also send us a fully completed accompanying slip with the measuring device to be checked. 

Download here  >>

Downloads Flow

LinkedIn Logo

Live Chat

This website is using Tidio, a chat platform that connects users with the customer support of E+E Elektronik Ges.m.b.H. We are collecting email addresses/names/phone numbers only with the consent of the users, in order to start the chat. The messages and data exchanged are stored within the Tidio application. For more information, please refer to their Privacy Policy. E+E Elektronik Ges.m.b.H is not making use of these messages or data other than to follow up on users’ registered issues or inquiries. Your personal data will be processed and transmitted in accordance with the General Data Protection Regulation (GDPR).