Reliable dew point measurement: what measurement data really demonstrate

Accuracy is only one aspect of measurement quality. For comprehensive process monitoring, long-term stability, repeatability, response time, behaviour after condensation and uninterrupted measurement output are equally important. Several test series for the TDS501 and EE355 and show how these characteristics can be substantiated with measurement data.

Long-term stability: observed over years

The long-term stability of the TDS501 was investigated separately using ten test units. The sensors were operated in Class 321 compressed air (DN50 compressed air line, 24-hour operation, typical dew point 60 °C (Td), pressure 8 bar and flow rate 100 Nm³/h) and repeatedly measured using a monitored dew point generator. The deviations at −70, −60 and −40 °C Td were evaluated over time as statistical envelopes. During the first six months, the determined envelopes at all three test points remained within ±2 °C Td. A 12-month projection was also derived from this trend. This projection likewise remains within the tolerance band. It is explicitly an extrapolation and not a long-term value that has already been measured.

Open link in lightbox
Fig. 1: TDS501 long-term stability at −70, −60 and −40 °C Td. The final point shows the 12-month projection derived from the first six months of measured data.

Long-term data are available for seven EE355 test units. Since March 2019, the transmitters have repeatedly been checked against a dew point reference at −50, −40 and −25 °C Td. The available time span differs between the test points: the data at −40 and −25 °C Td extend to May 2026, while the −50 °C Td series extends to February 2024.

At −50, −40 and −25 °C Td, all available measurement deviations remained within the documented tolerance band of ±2 °C Td. The behaviour is particularly consistent at −40 and −25 °C Td: between the first and the latest measurement, the deviation of each test unit changed by no more than 0.41 °C and 0.39 °C Td respectively.

Open link in lightbox
Fig.2: Maximum absolute deviation of the seven EE355 test units at each measurement date. The chart shows the dew point test points at −50, −40 and −25 °C Td.

Repeatability: comparable results under the same conditions

Repeatability was assessed using two TDS501 test units at −70, −65 and −50 °C Td. Each test point was stabilised for three hours; for the evaluation, the measured value immediately before the next change in the test point was used. The repeated measurements showed only very small deviations of less than ±0.1 °C Td at all three test points. The largest value was ±0.0845 °C Td.

Open link in lightbox
Fig. 3: Repeatability of two TDS501 test units. The chart shows the half-range of the repeated measurements at three dew point test points.

Response time: fast reaction to a humidity step

To simulate a humidity disturbance, a TDS501 was first operated for several hours in very dry air at approximately −74 °C Td. The test setup was then abruptly exposed to more humid air at around −20 °C Td. The sensor remained installed and in operation throughout the test. Evaluation of the step response gives a response time of approximately 15 seconds for T63 and approximately 30 seconds for T90. Major changes in dew point therefore become visible in the measurement signal at an early stage.

Open link in lightbox
Fig. 4: Dynamic response of the TDS501 to a dew point step from approximately −74 to −20 °C Td.

Condensation: still within specification after cyclic exposure

In compressed air applications, temporary condensation can occur in the event of faults or irregular operating conditions. To reproduce this stress, two TDS501 units underwent a condensation test based on EN 60068-2-30, a European standard specifying environmental testing procedures under high-humidity conditions. The test comprised twelve cycles of ten hours each at 93 ±3 % RH, with temperature changes between 15 ±3 °C and 55 ±3 °C. After the test, the characteristic curves were checked again at −70, −65, −60 and −40 °C Td. Both test units remained within the ±2 °C Td specification at all four test points.

The test demonstrates the measurement performance after cyclic condensation exposure. It should not be interpreted as a statement about continuous operation under condensation.

Open link in lightbox
Fig. 5: Measurement deviation of two TDS501 test units after the cyclic condensation test, based on EN 60068-2-30.

Uninterrupted measurements during auto-calibration

In addition to accuracy and stability, it is essential that the measured value remains continuously available during operation. In a dew point ramp from −70 to −65 °C Td at a rate of approximately 0.5 K/min, the TDS501 continued to provide readings throughout auto-calibration; its influence was barely visible in the signal trace.

A comparison over approximately 40 hours at around −56 °C Td illustrates the difference compared with a conventional auto-calibration method. Both sensors produced comparable average values of −55.9 and −55.8 °C Td. However, the conventional sensor held its last measured value during auto-calibration and then showed a jump of approximately 1.5 to 2 K. With the TDS501, there was no measurement interruption and the signal changes caused by auto-calibration were considerably smoother.

Open link in lightbox
Fig. 6: Comparison of the measurement signal during auto-calibration. The TDS501 provides continuous readings without signal freeze.

Measurement quality is more than an accuracy figure

The test series show that reliable dew point measurement is based on several characteristics: small deviation from the reference, stable readings over long periods, high repeatability, fast response to changes, robust behaviour after condensation exposure and continuous measurement output. Only the combination of these factors provides the basis for reliable process monitoring.

LinkedIn Logo

Live Chat

This website is using Tidio, a chat platform that connects users with the customer support of E+E Elektronik Ges.m.b.H. We are collecting email addresses/names/phone numbers only with the consent of the users, in order to start the chat. The messages and data exchanged are stored within the Tidio application. For more information, please refer to their Privacy Policy. E+E Elektronik Ges.m.b.H is not making use of these messages or data other than to follow up on users’ registered issues or inquiries. Your personal data will be processed and transmitted in accordance with the General Data Protection Regulation (GDPR).