The EE08 is optimized for accurate humidity and temperature measurement over a wide working range.
The probe has both analogue and digital outputs and is available with either cable or M12 connector. It is possible to choose between active and passive temperature measurement.
The probe is easily installed and dismounted thanks to the M12 thread. Low power consumption and short start-up time make the EE08 ideal for battery operated systems.
For meteorology and other outdoor applications, the EE08 can be used with the optional radiation shield.
With an optional confugration kit, it is possible to adjust the humidity and temperature reading and set up the analogue outputs.
Key Features
- Highest measurement accuracy
- Outstanding long term stability
- Wide working temperature range
- Compact and easily interchangeable
- Low power consumption
- Short start-up time
- Adjustment and outputs setup by the user
Typical Applications
- Meteorology, weather stations
- Hatcheries and incubators
- Climatic chambers
- Green houses, storage rooms
- Artificial snow machines
- Battery operated devices
Technical data
0...100% RH
Measuring range
-40...80 °C (-40...176 °F)±2% RH
Accuracy
±0.2 °C (±0.36 °F)analogue and E2 digital (2-wire)
Outputs4.5-15V or 7-30V DC
SupplyProduct inquiry
Send us your e-mail, we will do our best to help you!
Downloads
Datasheet EE08 Datasheet
(868.02 kb)EE-PCA Product Configuration Adapter
(790.77 kb)Scaling of the outputs
(252.78 kb)Accessories
(3893.46 kb)Humidity Calibration-Kit Datasheet
(428.89 kb)Manual EE08 Manual
(537.95 kb)E08 Manual Radiation Shield
(160.17 kb)Calibration Kit - User Guide
(1042.14 kb)Cleaning Instructions
(1054.21 kb)Support literature E2 Interface for humidity probes/modules
(569.13 kb)E2 Interface - specification
(223.84 kb)Utilising the E2 Interface
(663.63 kb)Software EE-PCS Product Configuration Software
(29086.25 kb)Related products
-
EE060 / EE061
Humidity and Temperature Probe with Analogue Output
Humidity and temperature probe with analogue outputs. Outstanding long term stability in polluted environment due to encapsulated electronics and robust sensor with E+E proprietary coating.
-
EE07
Humidity and Temperature Probe with Digital Output
Compact design, digital output and M12x1 connector facilitate easy installation and replacement of the EE07 humidity and temperature probe.
-
EE071
DISCONTINUED - Humidity and Temperature Probe with Modbus RTU
High accuracy, low power requirements and excellent protection against corrosion and pollution make this compact probe ideal for various indoor and outdoor applications.
Accessoires
-
HA010113
Metal grid filter
Construction
Material: polycarbonate
Filter: stainless steel wire mesh
Pores size: 30 µm
Length: 25mm (1")Features
For low mechanical stress
Low pollution level
For high RH / condensing environment
T range: -40...120°C (-40...248°F)
Response time t10/90: 15sTypical applications
Meteorology
Climate controlSuitable for Products:
-
HA011005
E2 to RS232 Converter
DESCRIPTION
- Converter for E2-interface to RS 232, incl. software für configuration, adjustement or test
- cable length: 2m (6.6ft)
Suitable for Products:
-
HA010703
Flange Coupling, 8 pins
-
HA010704
Flange Receptacle, 8 pins
-
HA010322
Connecting Cable, 8 pins - 1.5m
DESCRIPTION
- 8 pins
- M12x1 socket - flying leads
- shielded
- cable length: 1.5m (4.9ft)
Suitable for Products:
-
HA010323
Connecting Cable, 8 pins - 3m
DESCRIPTION
- 8 pins
- M12x1 socket - flying leads
- shielded
- cable length: 3m (9.8ft)
Suitable for Products:
-
HA010324
Connecting Cable, 8 pins - 5m
DESCRIPTION
- 8 pins
- M12x1 socket - flying leads
- shielded
- cable length: 5m (16.4ft)
Suitable for Products:
-
HA010325
Connecting Cable, 8 pins - 10m
DESCRIPTION
- 8 pins
- M12x1 socket - flying leads
- shielded
- cable length: 10m (32.8ft)
Suitable for Products:
FAQ
-
Which is the maximum cable length of E2 interface?
The maximum cable length depends on the product, clock speed and pull-up resistor, and can be up to 10m. For detail please see E2 interface for humidity probes/modules.
-
How many E+E devices can be connected on one bus to a controller with E2 interface?
Only one E+E device can be connected to the E2 interface of a controller.
-
What is the difference between a 2 wire and a 3 wire current output?
Transmitter with "two-wires" technology receive the power from the process, and the signal is carry on return wire (closed current loop). With technology "three-wires" the power supply is separate from current output: 2 wires are the power supply and third carry the signal.