High-End Moisture in Oil Transmitter with multifunctional Display

  • Service
  • Moisture in Oil

EE360

EE360 is dedicated for reliable monitoring of lubrication, hydraulic and insulation oils as well as diesel fuel. In addition to highly accurate measurement of water activity (aw) and temperature (T), EE360 calculates the absolute water content (x) in ppm.

The probe can be employed up to 180 °C (356 °F) and 20 bar (290 psi) and is available with either ISO or NPT slide fitting, which allows for variable immersion depth. Using the optional ball valve, the probe can be mounted or removed even without process interruption.

The design of the enclosure facilitates easy mounting and maintenance. EE360 is available with IP65 polycarbonate or stainless steel enclosure. The measured values are available on two analogue outputs and on the optional digital interface RS485 with Modbus RTU or Ethernet with Modbus TCP. An optional relays module can be used for alarms and process control.

The state of the art TFT colour display can show all measurands simultaneously and offers extensive error diagnostics. The integrated data logging function saves all measured data in the internal memory. The logged data can be displayed in a graph directly on the device or easily downloaded via USB interface.

The EE360 configuration and adjustment can be performed either directly on the device via display and push buttons or with the free  PCS10 product configuration software using the USB service interface.

Typical applications

Monitoring of

  • transformer oil
  • lubrication oil
  • hydraulic oil
  • diesel fuel

EXTEND YOUR WARRANTY UP TO 5 YEARS WITH THE NEW 
E+E CALIBRATION SERVICE PACKAGES

Our calibration service packages ensure that you are on the safe side when it comes to the accuracy and vailability of your measuring devices. Choose the right service package for your E+E product and benefit rom professional device service, calculable costs and excellent measuring accuracy in the long-term. Here you can find more detailed information.

We are always available to answer any questions you may have! Inquire now
  • UL listed versions
  • Measurands: water activity (aw), temperature (T) and water content (x) in ppm
  • TFT colour display with push buttons
  • Two analogue outputs
  • Digital interface RS485 with Modbus RTU or Ethernet with Modbus TCP
  • Integrated data logging for 20.000 values
  • Error self diagnosis shown on display or via error LEDs
  • Auditive and visual error warnings
  • Remote probe up to 20m (65.6ft)
  • Rugged polycarbonate IP65 or stainless steel enclosure
  • Modular enclosure concept for easy mounting and service
  • Pressure tight probe up to 20 bar (290 psi)
  • Pluggable probe for fast exchange and easy installation
  • Two freely configurable alarm outputs
  • G ½ ISO or ½ NPT process connection
  • Optional ball valve for probe mounting and removal without process interuption
  • Power supply module for 100...240 V AC
  • Easy field adjustment and configuration via display or free software
Measuring range
0...1 aw0...100,000 ppm-40...180 °C (-40...356 °F)
Accuracy - °measured value
±(0.015 + 1.5%*mv°) aw±0.2 °C
Digital interface
RS485 with Modbus RTUEthernet with Modbus TCP
Outputs
0-1/5/10V, 0/4-20mA
Supply
8...35V DC12...30 V AC
You want to know more about our products? Inquire now

FAQ

Use the RS485 kit for retrofitting (order number HA010605), see the data sheet and user manual of your specific E+E product.

The device indicates an error according to NAMUR NE43. For details, please see the user manual of your specific E+E product.

The device has one GND connection for the power supply, an analogue output GND and an interface GND, see connection diagram in the user manual. The analogue output GND and the interface GND carry only the current output resp. interface current, which results in a more stable signal compared to a mix GND connection.

Press 2x ► | 5x ◄ | 1x ▲ | 3x ◄ | 1x ▼ | 1x ► then select the "Configuration menu" and press ENTER. Scroll down to “Diagnosis”, press ENTER and scroll down to “Password protection” here you can see the previously set password.

Connect the device via the USB service interface to a PC running PCS10 Product Configuration Software and read out the communication setting.

Alternative for device with display: the display configuration menu has a section for Modbus settings (RS485, Modbus RTU) or IP settings (Ethernet, Modbus TCP) where you can see the actual setting or change it.

Connect the device via the USB service interface to a PC running PCS10 Product Configuration Software and see at the section "Log data download" for the data memory status.

Remote probe types can be retrofitted with an Ethernet interface using HA010606, see data sheet and operation manual of your specific E+E product.

The device requires a firmware update. Connect the device via the USB service interface to a PC running PCS10 Product Configuration Software and confirm the update to version 1.20 or greater.

The measured data logged by the device can only be downloaded via the USB service interface to a PC running the PCS10 Product Configuration Software.

The pluggable remote probe has a label with the serial number close to the connector. This must be the same as the serial number of the basis device. Refer to the operation manual further details.

Calibration for a specific oil is a laboratory procedure for determining the saturation curve of the oil. The saturation curve is described by two parameters (A and B). These parameters are used for setting up the moisture in oil sensor for correct calculation of the water content (ppm) out of the measured temperature and water activity (aw).

No, coating is not available for the E+E MIO sensors.

Sensor replacement is possible for the EE360 models with the ordering code including PC4 (pluggable probe). Please see the operation manual or detail and spare part code ordering guide.

Please see the cleaning instructions.

The standard scope of supply of E+E MIO sensors includes an inspection certificate according DIN EN 10204 - 3.1. For details on calibration certificates please see  our white paper "Calibration and Traceability in Measuring Technology".

No, mineral transformer oils from different manufacturers are quite similar. A and B parameters are the same.

We are always available to answer any questions you may have! Inquire now

Downloads EE360

You want to know more about our products?

Explore E+E technology and high-end sensor calibration. Our global team helps you find a custom solution for your production.

Matthew Nemeth | Managing Director E+E Elektronik Corporation USA
Matthew Nemeth | Managing Director E+E Elektronik Corporation USA
LinkedIn Logo

Live Chat

This website is using Tidio, a chat platform that connects users with the customer support of E+E Elektronik Ges.m.b.H. We are collecting email addresses/names/phone numbers only with the consent of the users, in order to start the chat. The messages and data exchanged are stored within the Tidio application. For more information, please refer to their Privacy Policy. E+E Elektronik Ges.m.b.H is not making use of these messages or data other than to follow up on users’ registered issues or inquiries. Your personal data will be processed and transmitted in accordance with the General Data Protection Regulation (GDPR).